Application Note

Assess virus neutralization with a rapid, HTS-friendly assay

  • Rapid, easy assay setup with an ELISA workflow
  • No BSL-3 or cell culture required
  • Results consistent with data from plaque reduction neutralization test

Download PDF

Cathy Olsen, PhD | Sr. Applications Scientist | Molecular Devices

Introduction

The worldwide COVID-19 pandemic caused by the SARS-CoV-2 virus has necessitated the fast-tracked development of many research tools for understanding this virus’s pathogenesis, as well as vaccine discovery and development. Assays for monitoring the immune response to infection by and vaccination against the virus are important to COVID-19 research. As neutralizing antibodies are key biomarkers of immune response and vaccine efficacy, levels of neutralizing antibodies in patient serum samples is an important parameter to be able to monitor efficiently.1

Identification of neutralizing antibodies in human serum has commonly been performed using traditional methods like the plaque reduction neutralization (PRNT) assay. However, because PRNT uses live, infectious virus added to target cells, it requires a BSL-3 laboratory and laborious cell culture techniques. Analysis of the resulting data is time consuming and not amenable to automation. Alternatively, the pseudovirus neutralization test (pVNT) method can be done in a BSL-2 lab using a modified form of the virus, but it requires cell culture and fluorescence imaging to measure neutralizing activity, so it is also not suited to high-throughput screening of samples.2

The GenScript cPass SARS-CoV-2 Neutralization Antibody Detection Kit is a surrogate virus neutralization test (sVNT) developed to overcome the hurdles of more conventional virus neutralization assays. It employs an enzyme-lnked immunosorbent assay (ELISA) format as outlined in Figure 1 and graphically depicted in Figure 2.

An HRP-conjugated fragment of the receptor-binding domain (RBD) of the spike protein is combined with a patient serum sample, which may contain neutralizing antibodies. This mixture is then added to the ELISA plate, which is coated with ACE2 receptor. If antibodies in the sample have neutralizing activity, binding of the HRP-RBD to ACE2 will be disrupted, and subsequent washing will remove the HRP-RBD. The resulting signal, measured using an absorbance microplate reader, will be lower in the presence of neutralizing antibodies, and comparison of the signal to that of assay controls enables assessment of the neutralizing activity present in the test samples.3

Here, results for the sVNT kit detected with SpectraMax® microplate readers and analyzed using SoftMax® Pro Software are presented. These results correlate closely with results previously generated using the traditional PVNT method, while only requiring a small fraction of the time and a less restrictive laboratory biosafety level.

Comparison of PRNT and sVNT kit workflows

Figure 1. Comparison of PRNT and sVNT kit workflows. The sVNT assay can be completed in one hour in a BSL-2 lab, while PRNT requires live SARS-CoV-2 virus and cumbersome cell culture techniques, taking more than two days at a BSL-3 lab.

SARS-CoV-2 sVNT method

Figure 2. SARS-CoV-2 sVNT method. The assay uses an ELISA format that enables detection of disruption of RBD binding to ACE2 by neutralizing antibodies in a test sample.

Materials

Methods

All kit components and test samples were allowed to reach room temperature prior to assay set up.

Kit reagents requiring dilution were prepared as follows:

Test samples and assay controls were mixed with HRPconjugated RBD and incubated as follows. Test samples, along with positive and negative controls, were diluted 1:10 with sample dilution buffer, e.g., 12 μL test sample + 108 μL buffer. Each diluted sample and control was mixed in separate tubes with an equal volume of HRP-RBD working solution, e.g., 120 μL HRP-RBD working solution + 120 μL diluted test sample or control. Mixtures were incubated at 37°C for 15 minutes.

100 μL of positive control mixture, negative control mixture, and sample mixtures were added to duplicate wells of the assay strips. The plate was then sealed and incubated at 37°C for 15 minutes.

The MultiWash+ plate washer was used to wash all wells four times with 300 μL of 1X wash solution per well, followed by a final crosswise aspiration. 100 μL of TMB Solution was then added to each well, and the plate was sealed and incubated in the dark at room temperature for 15 minutes.

Cutoff
Result
Interpretation
≥ 30%
Positive
SARS-CoV-2 neutralizing antibody detected
< 30%
Negative
No detectable SARS-CoV-2 neutralizing antibody

Table 1. Percent inhibition cutoff values used to interpret assay results.

After incubation with TMB substrate, 50 μL of stop solution was added to each well, and absorbance at 450 nm was read immediately on each microplate reader.

Validity of the test results was assessed by determining whether the negative control OD450 values were above 1.0, and the positive control values were below 0.3. Failure to meet these criteria would invalidate the results of the test and require it to be repeated.

Percent inhibition for each test sample was calculated using the following formula:

Inhibition = (1 - OD value of Sample/OD value of negative control) x 100%

The cutoff values in Table 1 were used to interpret sample results as positive or negative, interpreted as the presence or absence of detectable SARS-CoV-2 neutralizing antibody. Cutoff values shown in the table are from the SARS-CoV-2 sVNT kit manual. For different geographic locations or ethnic backgrounds, users are directed to set up their own cutoff values based on representative patient serum panels.

Results

Quality of the assay results was assessed using the negative and positive controls. As shown in Table 2, the OD450 values for the negative control wells were above 1.0, and OD450 values for the positive control wells were below 0.3, as specified in the quality control requirements for the sVNT kit.

The OD450 values of the samples and of the negative control were used to calculate percent inhibition, as described in Methods. Percent inhibition values at or above 30% were interpreted as positive for the presence of SARS-CoV-2 neutralizing antibody, and values below 30% were interpreted as negative, based on the general cutoff values specified in the kit manual. Each of the three samples that had tested negative for neutralizing activity via PRNT also tested negative with the sVNT assay. Likewise, ten samples that tested positive via PRNT also tested positive via sVNT (Table 3).

Control

OD

450

Average
Std Dev
%CV
Negative
2.066 2.148
2.107
0.058
2.7
Positive
0.063 0.061
0.062
0.002
2.8

Table 2. Negative and positive control OD450 values were above 1.0 and below 0.3, respectively, validating the assay results. Controls were run in duplicate, with %CV values below 3.0 demonstrating reproducibility.

Plaque reduction

neutralization test

GenScript sVNT assay
Sample
PRNT90

Average OD

450

%

Inhibition

Result
N-1
<20
1.991
5.5
Negative
N-2
<10
1.869
11.3
Negative
N-3
<20
1.590
24.6
Negative
P-1
160
0.112
94.7
Positive
P-2
80
0.061
97.1
Positive
P-3
80
0.096
95.4
Positive
P-4
160
0.091
95.7
Positive
P-5
320
0.084
96.0
Positive
P-6
320
0.141
93.3
Positive
P-7
160
0.105
95.0
Positive
P-8
640
0.101
95.2
Positive
P-9
640
0.149
92.9
Positive
P-10
160
0.225
89.3
Positive

Table 3. Percent inhibition and positive vs. negative results for three samples that previously tested negative, and ten samples that tested positive, with a plaque reduction neutralization test (PRNT). Results shown were obtained using the SpectraMax ABS Plus reader; identical results (positive vs. negative) were obtained with the other SpectraMax readers.

Conclusion

The GenScript cPass SARS-CoV-2 Neutralization Antibody Detection Kit data obtained using a set of patient serum samples matched results generated with the more traditional and laborious PRNT method. The sVNT kit offers easy-to-use reagents and a compact ELISA workflow that could be completed in about an hour. The one required wash step was automated using the MultiWash+ plate washer to save valuable time. SpectraMax readers with SoftMax Pro Software generate data that can be analyzed using an assay-specific protocol with formulas configured to apply the required calculations and output results automatically.

References

  1. Lau EHY, Tsang OTY, Hui DSC, Kwan MYW, Chan W, Chiu SS, Ko RLW, Chan KH, Cheng SMS, Perera RAPM, Cowling BJ, Poon LLM, and Peiris M. Neutralizing antibody titres in SARS-CoV-2 infections. Nature Communications 12:63 (2021).
  2. Sahin U, Muik A […] and Tureci O. COVID-19 vaccine BNT162b1 elicits human antibody and TH1 T cell responses. Nature 586, 594–599 (2020).
  3. SARS-CoV-2 surrogate Virus Neutralization Test (sVNT) Kit Manual (cat. #L00847).

Download PDF